Back to Catalogue

   Back to Water Category

   Reference Normes



Codes indicated with numbers:

11xxx.1 supplied with G 3/8" coupling

11xxx.2 supplied with G 1/2" coupling

11xxx.3 supplied with G 3/4" coupling

1. To require chemical nickel plating add NK to the article's code
ex. 11001.1NK

2. To require ceramics water head valve (1/4 turn) add C to the article's code
ex. 11001.1C




Water taps can be supplied with oil bath head valve and sealing in EPDM or with ¼ turn ceramics head valve.

Nozzle with possibility of disassembly in compliance with DIN 12898 regulation.

Handles in compliance with EN13792:2000 norm, manufactured in ABS;
acid resistant material.

Threads in compliance with IS0228/1 regulation, B tolerance class.

Taps are built following the DIN12918 norm.


FAR can furthermore supply for demineralised water all the series with an internal chemical nickel-plating with a tickness of 17 microms, to have the desired article with this treatment the customer just needs to add to the article the NK abbreviation.

For ultra pure waters FAR manufactures PP and PVDF taps with recirculation.
Maximum working pressure 10 Bar.


Hereunder are shown the charts of water taps output; these results were acquired in the following conditions: operational pressure 10 Bar, temperature included between 10° and 25° and with a range of measurement uncertainty of +/- 2%.

The output curves "Qw" as a function of the supply pressure "P" hereunder indicated were interpolated previously, using the method of square minimums, in order to reach the equation of the characteristic curve Qw = f x (p) of the general kind Q = k x p raised to the nth power.




Tap Model
Body
Type
Head Valve
Type
Characteristic Equation
11000.1 with nozzle
straight
normal
Qw = 7,929019 . P 0,4912054
11000.1 without nozzle
straight
normal
Qw = 11,21703 . P 0,4865208
11001.1 with nozzle
90°
normal
Qw = 9,539138 . P 0,4885823
11001.1 without nozzle
90°
normal
Qw = 25,53553 . P 0,4698563
11001.1 with nozzle
90°
ceramic
Qw = 7,678724 . P 0,4925607
11001.1 without nozzle
90°
ceramic
Qw = 11,3417 . P 0,4916656
11003.1 with nozzle
45°
normal
Qw = 9,715805 . P 0,486955
11003.1 without nozzle
45°
normal
Qw = 27,05927 . P 0,4867322
11310.0  
//
normal
Qw = 8,282887 . P 0,3923514
11310.0  
//
ceramic
Qw = 6,608737 . P 0,4013889

 Characteristic Equation: [with Qw in 1/min. and P in Bar]




Hydraulic characteristic of water tap for laboratory, art. 11000.1
with normal head valve
zoom
1 . with nozzle
2 . without nozzle


Hydraulic characteristicof water tap for laboratory, art. 11001.1
with normal head valve
zoom
1 . with nozzle
2 . without nozzle


Hydraulic characteristic of water tap for laboratory, art. 11000.1
with ceramic head valve
zoom
1 . with nozzle
2 . without nozzle


Hydraulic characteristic of water tap for laboratory, art. 11003.1
with normal head valve
zoom
1 . with nozzle
2 . without nozzle


Hydraulic characteristic of water tap for laboratory, art. 11310.1
with normal head valve
zoom
1 . with nozzle
2 . without nozzle


Hydraulic characteristic of water tap for laboratory, art. 11310.0
with ceramic head valve
zoom
1 . with nozzle
2 . without nozzle